Haskris Water Chiller for Bruker AXS System 230V

来自 Haskris
可用性:
只1左边!
库存量单位 11114
节省49% 节省49%
原价 $6,500.00
原价 $6,500.00 - 原价 $6,500.00
原价 $6,500.00
现价 $3,300.00
$3,300.00 - $3,300.00
现价 $3,300.00

Haskris R175-BRUKERAXS-ABCHKO High-Capacity Recirculating Water Chiller for Bruker AXS X-Ray Systems. An OEM-customized, heavy-duty closed-loop refrigerated liquid chiller specifically engineered to provide high-stability, continuous-duty cooling for Bruker AXS analytical X-ray diffraction (XRD), X-ray fluorescence (XRF), and single-crystal crystallography instrumentation.

Used; excellent condition; see video at this link.

  • MPN / OEM Catalog Configuration: R175-BRUKERAXS-ABCHKO (Haskris R175 Series OEM Custom Build for Bruker AXS)
  • Manufacturer & Line: Haskris Company / R-Series Heavy-Duty Analytical & Medical Chiller Systems
  • Target System Compatibility: Dedicated OEM cooling solution tailored to meet the hydraulic pressure, flow rate, and thermal stability tolerances of Bruker AXS X-ray generators and tubes (including D8 ADVANCE, D8 DISCOVER, D2/D4 ENDEAVOR, S8 TIGER, and SMART APEX/Quest platforms)
  • Nominal Cooling Capacity: Approx. 1.75 Tons (~21,000 BTU/hr / ~6,150 W at 20°C fluid supply temperature), capable of dissipating continuous high-wattage X-ray anode heat loads
  • Operating Temperature Range & Precision: Closed-loop PID microprocessor temperature regulation spanning +5°C to +35°C (typically set at 18°C–20°C for X-ray tubes) with tight thermal stability within ±0.1°C to ±0.2°C to prevent thermal drift in goniometer alignment and detector electronics
  • Circulation Pump & Hydraulic Performance: Heavy-duty positive displacement (rotary vane) or multi-stage centrifugal pump delivering high-head fluid circulation (typically 3.0 to 6.0 gpm at regulated pressures of 50 to 80+ psi) with an integrated adjustable pressure bypass valve
  • Wetted Path & Coolant Purity: Corrosion-resistant, non-ferrous wetted path (stainless steel reservoir and heat exchanger, brass/copper/polymer plumbing) compatible with distilled water, low-conductivity deionized (DI) water, or approved inhibited glycol mixtures required by high-voltage X-ray tube assemblies
  • Refrigeration & Condenser Architecture: Hermetically sealed industrial refrigeration compressor using CFC/HCFC-free refrigerant; features a high-efficiency air-cooled fin-and-tube condenser with louvered front airflow intake (or water-cooled condenser variant depending on facility hookup)
  • Instrumentation, Interlocks & Safety Suite: Front-mounted digital temperature controller and pressure readout, external coolant level sight glass, low-flow / low-level safety interlocks, remote start/stop and fault relay contacts for direct hardware interlock integration with the Bruker AXS X-ray generator safety circuit
  • Electrical Specifications: 208/230 VAC, 60 Hz (single-phase or three-phase facility circuit configuration per OEM spec) with integrated circuit breaker protection
  • Chassis & Form Factor: Heavy-gauge powder-coated steel floor-standing cabinet with removable side service panels for straightforward preventive maintenance, filter cleaning, and reservoir servicing
  • Primary Application: Continuous heat removal from high-power sealed X-ray tubes, rotating anode generators, high-voltage power supplies, and solid-state X-ray detectors in materials science, pharmaceutical polymorphism, geological, and industrial QA/QC laboratories

请注意,Life Sciences Trading 不是本店所列产品的厂家授权经销商。这些产品是从领先的生命科学公司收购的剩余产品。